The testo Saveris 2 WiFi data logger system is the easy, versatile and reliable solution for the measurement and monitoring of temperature and door contact values in your storage, refrigerated and work rooms. Installing the system is child's play and can be carried out either via a browser or with the testo Saveris 2 App.
Temperature monitoring with testo Saveris 2
The temperature monitoring system consists of at least one WiFi data logger and a Cloud access (online data store). As a component of this data logger system, the testo Saveris 2-T2 data logger with up to 2 connectable NTC temperature probes or door contacts stores all measured temperature values and door contacts and transmits them directly by wireless LAN to the Testo Cloud. You can use the alarm function to receive immediate warnings by e-mail or SMS (optional) when limit values are violated.
An overview of the benefits
The Testo Cloud is the core operating element for setting up your testo Saveris 2 system. Here you can configure your WiFi data loggers, set limit value alarms and analyze your measurement data. You can access all measuring values and analysis functions anywhere and at any time with an internet-capable smartphone, tablet or PC. You must first register at www.saveris.net to have access to the Testo Cloud.
|
Advanced |
Measuring cycle |
1 min. ... 24 h |
Communication cycle |
1 min. ... 24 h |
Data storage |
Max. 2 years |
Reports |
Manual (.pdf/.csv)
Automatic (.pdf/.csv) |
Number of users per account |
10 |
Number of WiFi data loggers
per account |
Unlimited |
Alarm options |
• Upper/lower alarm limits
• Alarm delay
• Time control of alarms |
System notifications |
• Notification of low battery
• Radio link interrupted
• Power supply interrupted |
E-mail alarm |
Yes |
SMS alarm |
• Including 25 SMS per logger and year
• Option of purchasing additional SMS packages |
Measurement data export (API interface) |
Yes |
WiFi data logger: advantages and functions
Current measuring values, limit value violations and the remaining battery life can be seen in the data logger's display. The standard batteries (AA) have a lifetime of 12 months and can be replaced at any time by users themselves. The internal data memory of the WiFi data logger is 2 x 10,000 readings and the housing has IP65 protection. The testo Saveris 2-T2 is tested by TÜV Süd according to EN 12830.
Hint:
In order to use the testo Saveris 2-T2 WiFi data logger, you also require at least one connectable probe (optional).
- testo Saveris 2-T2 WiFi data logger
- Micro USB cable
- Wall bracket with lock
- Batteries (4 x AA AlMn Mignon batteries)
- Short instructions
- Calibration protocol
Hint:
In order to use the testo Saveris 2-T2 WiFi data logger, you also require at least one connectable probe (optional).