1. Home
    2. testo 330i - flue gas CO/NO

    testo 330i - flue gas analyzer with NO sensor

    0632 3000 72
    SAR 6 975
  1. On-site digital creation of reports with the testo 330i App
  2. Operation independent of measuring location – readings are always available when you need
  3. testoFix probe holder (optional) – holds the probe in the centre of flow at all times
  4. Particularly robust, completely enclosed plastic housing
  5. Proven technology meets revolutionary operation. The testo 330i flue gas analyzer was specifically developed to help you to overcome all the obstacles you meet when servicing and commissioning heating systems.  
    The testo 330i flue gas analyzer is an advancement of the testo 330 LL flue gas analyzer. Its proven measuring technology is once again at the heart of the testo 330i. This means that you can rely on highly accurate results when measuring flue gas, draught and O2 supply air. The testo 330i also has the following impressive features:
    • User-replaceable long-life sensors with a service life of up to 6 years
    • Successful TÜV inspection according to the 1st German Federal Immission Control Ordinance (BImSchV) and EN 50379, Parts 1-3
    • ntegrated draught and gas zeroing. This means that the probe does not have to be removed from the flue gas duct for the zeroing
    • Fresh air dilution up to 30,000 ppm upon exceeding a CO threshold value
    • NO sensor for combined heat and power plants
    The new testoFix probe holder is available as an accessory, so the flue gas probe is guaranteed not to lose the centre of flow once it has been found. This can be attached to the pipe quickly and easily. It holds the probe in the centre of flow and also serves as a fall-proof suspension system for the testo 330i flue gas analyzer. Incidentally, the hose between probe and measuring instrument is lighter and more flexible than before.

    The future of flue gas measuring technology

    The testo 330i no longer has a display. How do you make the correct adjustments to the flue gas analyzer? Where can you view the readings? What about the creation of measurement protocols? Via the testo 330i App on your smartphone. The readings are transferred from the testo 330i flue gas analyzer to the App via Bluetooth. So you always have the relevant values to hand whenever you need them. This is particularly useful when faced with larger systems or awkward measuring points as there is no need to look around corners, which can be annoying. The measurement parameters can also be set intuitively, via the App. Once you have completed the measurement, you can add comments or photos of the system to the results on site, and send them to your customer or the office via e-mail.

    Requirement: iOS 7.1 / Android 4.3 or newer. Mobile end device with Bluetooth 4.0.

    testo 330i flue gas analyzer with long-life gas sensors, 3 slots and integrated draught and gas zeroing, including O2/CO measuring cell, NO sensor, rechargeable battery and calibration protocol.  

    General technical data

    Weight
    720 g (excluding battery)
    Dimensions
    270 x 160 x 57 mm
    Operating temperature
    -5 to +45 °C
    Power supply
    Rech. batt. block 3.7 V / 2.6 Ah; Mains unit 6 V / 1.2 A (optional)
    System requirements
    requires iOS 7.1 or newer; requires Android 4.3 or newer; requires mobile end device with Bluetooth 4.0
    Maximum memory
    500000 readings
    Storage temperature
    -20 to +50 °C

    Differential Pressure - Piezoresistive

    Measuring range
    0 to 300 hPa
    Accuracy
    ±1.5 % of mv (Remaining Range)
    ±1 % of mv (+50.1 to +100.0 hPa)
    ±0.5 hPa (0.0 to +50.0 hPa)
    Resolution
    0.1 hPa

    Flue gas Oâ‚‚

    Measuring range
    0 to 21 Vol.%
    Accuracy
    ±0.2 Vol.%
    Resolution
    0.1 Vol.%

    Accessories for probes

    Combustion air / temperature

    Probes

  6. Probes
    Flue gas probe 300mm, 8mm
    Flue gas probe 300 mm, diameter 8 mm
      :
      0600 9781
      SAR 1 121
    • Probes
      Flue gas probe 180mm, 6mm
      Flue gas probe 180 mm, diameter 6 mm
        :
        0600 9782
        SAR 1 058
      • Probes
        Flue gas probe 180mm, 8mm
        Flue gas probe 180 mm, diameter 8 mm
          :
          0600 9780
          SAR 1 058
        • Probes
          Flue gas probe 300mm, 6mm
          Flue gas probe 300 mm, diameter 6 mm
            :
            0600 9783
            SAR 1 121
          • Product sets

          • Product sets
            testo 330i – flue gas set 
              :
              0563 3000 70
              SAR 8 341
            • Product sets
               
                :
                0563 3000 71
                SAR 8 744
              • Accessories

                Accessories for measuring instrument

              • Accessories for measuring instrument
                Smoke tester with oil and soot sheet, for measuring soot in flue gas, excl. cone (part no. 0554 9010)
                  :
                  0554 0307
                  SAR 597
                • Accessories for measuring instrument
                  Spare particle filter
                  Spare dirt filters - for probe handle (10 off)
                    :
                    0554 3385
                    SAR 119
                  • O2 supply air measurement

                    In balanced flue systems, is there perhaps a leak in the flue gas system? Use the testo 330i flue gas analyzer to find out: simply measure the O2 content of the combustion air in the dual wall clearance. This way, you can ensure that no flue gas is getting into the combustion air supply. This prevents damage due to condensate or impairment of the combustion process.

                    Differential pressure measurement

                    Measure the gas flow and gas resting pressure on gas boilers to ensure that the system can be commissioned properly. You can also use the data specified by the manufacturer for the nozzle pressure to tailor the appliance output to the actual heat requirement.

                    Flue gas analysis and flue gas measurement

                    The testo 330i flue gas analyzer features two user-replaceable long-life sensors for measuring O2 and CO as well as an NO sensor. Flue gas and draught can be measured in parallel using the testo 330i – so you can see all the key values simultaneously in the testo 330i App on your smartphone or tablet.

                    • Testo ZIV driver in the 2000 version. The Testo ZIV driver is used to connect the testo 320 and testo 330 measuring instruments to an application program (sweeping district administration program) according to the interface Version 2.0 defined by the Zentralverband des Schornsteinfegerhandwerks (Central Association of Chimney Sweeps, ZIV). Please check with the manufacturer of your application program as to whether this interface is supported. If Microsoft .NET Framework 4.0 has not been installed on the computer, it must be downloaded from the Microsoft website and installed on the system.
                    • The Testo ZIV driver is used to connect the testo 300, testo 320 and testo 330 measuring instruments to an application program (sweeping district administration program) according to the interface defined by the Zentralverband des Schornsteinfegerhandwerks (ZIV, Central Association of Chimney Sweeps) in version 1.0 of 01. August 2012, in version 2.0 of 13. February 2017 as well as version 3.0 from 02. July 2021. Please check with the manufacturer of your application program as to whether this interface is supported.
                    • Firmware testo 330i(v2.11, 3.56 MB)
                      If the firmware update does not start under Windows 8.1 or Windows 10, a new bootloader must be installed on the measuring device once.
                      A description and all necessary files can be found under the search term: Update-Kit / Bootloader
                    • Update-Kit / Bootloader(V1.22, 1.24 MB)
                      (testo 330 LL | testo 330i | testo 350 Control Unit + Analysis Box | testo 320)
                      If the firmware update does not start under Windows 8.1 or Windows 10, a new bootloader must be installed on the measuring device once.