1. Home
    2. testo 330-2 LL

    testo 330-2 LL

    0632 3307 70
    SAR 5 766
  1. testo 330-2 LL flue gas analyzer with long-life gas sensors and built-in draugh and gas zeroing; BLUETOOTH and with H2-compensated CO-sensor; incl. rech. battery and calibration protocol; with graphic display; version 2010
    Thanks to the new instrument functions, the Testo flue gas analyzers of the new testo 330-LL series now offer you even more professional and reliable support. These are the main features:
    • High-resolution colour display for the graphic representation of your measurement data
    • Extended measurement menus – such as gas pipe testing – for a comprehensive analysis of your heating system
    • Logger function for easy long-term recording of the measurement curve

    General technical data

    Weight
    600 g (without rechargeable battery)
    Dimensions
    270 x 90 x 65 mm
    Operating temperature
    -5 to +45 °C
    Display size
    240 x 320 pixels
    Display function
    Colour graphic display
    Power supply
    Rechargeable battery pack 3.7 V / 2.6 Ah; Mains unit 6 V / 1.2 A
    Maximum memory
    500,000 readings
    Storage temperature
    -20 to +50 °C

    Differential Pressure - Piezoresistive

    Measuring range
    ±10000 Pa
    Accuracy
    ±3 % of mv (10 to 10000 Pa) plus ±1 Digit
    ±0.3 Pa (0 to 9.99 Pa) plus ±1 Digit

    Flue gas O₂

    Measuring range
    0 to 21 Vol.%
    Accuracy
    ±0.2 Vol.%
    Resolution
    0.1 Vol.%
    Reaction time t₉₀
    < 20 s

    Sets

    Ambient CO measurement in the heated environment

    Carbon monoxide (CO) is a colourless, odourless and taste-free gas, but also poisonous. It is produced during the incomplete combustion of substances containing carbon (oil, gas, and solid fuels, etc.). If CO manages to get into the bloodstream through the lungs, it combines with haemoglobin thus preventing oxygen from being transported in the blood; this in turn will result in death through suffocation. This is why it is necessary to regularly check CO emissions at the combustion points of heating systems and the surroundings.

    testo 330-1/-2 LL

    Measuring pressure on burners (nozzle pressure, gas flow pressure, etc.)

    Standard readings taken during services of domestic heating systems include checking the gas pressure on the burners. This involves measuring the gas flow pressure and gas resting pressure. The flow pressure, also called supplied pressure, refers to the gas pressure of the flowing gas and resting pressure of the static gas. If the flow pressure for gas boilers is slightly outside the 18 to 25 mbar range, adjustments must not be made and the boiler must not be put into operation. If put into operation nonetheless, the burner will not be able to function properly, and explosions will occur when setting the flame and ultimately malfunctions; the burner will therefore fail and the heating system will shut down.

    Measuring the flue gas parameters of the burner (CO, O2, and temperature, etc.)

    The flue gas measurement for a heating system helps to establish the pollutants released with the flue gas (e.g. carbon monoxide CO or carbon dioxide CO2) and the heating energy lost with the warm flue gas. In some countries, flue gas measurement is a legal requirement. It primarily has two objectives:

    1. Ensuring the atmosphere is contaminated as little as possible by pollutants; and
    2. energy is used as efficiently as possible.

    Stipulated pollutant quantities per flue gas volume and energy losses must never be exceeded.

    Measurement in terms of results required by law takes place during standard operation (every performance primarily using the appliance). Using a Lambda probe (single hole or multi-hole probe), the measurement is taken at the centre of flow in the connecting pipe (in the centre of the pipe cross-section, not at the edge) between the boiler and chimney/flue. The measured values are recorded by the flue gas analyzer and can be logged either for print out or transfer to a PC at a later stage.

    Measurement is taken by the installer at commissioning, and if necessary four weeks later by the flue gas inspector/chimney sweep, and then at regular intervals by the authorised service engineer.

    testo 330-1/-2 LL

    Measuring temperatures on radiators

    When measuring temperatures on radiators, the flow and return temperature are recorded in particular and assessed by the tradesman. The flow temperature is defined as the temperature of the thermal transfer medium (e.g. water) that the system is supplied with. The temperature of the medium flowing out of the system is accordingly called the return temperature. To prevent losses within the heat distribution system and achieve a better level of efficiency spot recording of flow and return temperatures are necessary. Implementation of relevant measures ultimately leads to hydraulic adjustment on the basis of knowledge about the flow and return temperatures. This defines a procedure with which every radiator or heating circuit of a flat radiator within a heating system is supplied at a set flow temperature with the precise amount of heat needed to achieve the ambient temperature required for the individual rooms. Flawed operating conditions will result in considerable excess consumption of electricity and heating energy. The German Energy Saving Regulation (EnEV) therefore requires hydraulic adjustment for systems being set up or overhauled for this very reason.

    • Testo ZIV driver in the 2000 version. The Testo ZIV driver is used to connect the testo 320 and testo 330 measuring instruments to an application program (sweeping district administration program) according to the interface Version 2.0 defined by the Zentralverband des Schornsteinfegerhandwerks (Central Association of Chimney Sweeps, ZIV). Please check with the manufacturer of your application program as to whether this interface is supported. If Microsoft .NET Framework 4.0 has not been installed on the computer, it must be downloaded from the Microsoft website and installed on the system.
    • The Testo ZIV driver is used to connect the testo 300, testo 320 and testo 330 measuring instruments to an application program (sweeping district administration program) according to the interface defined by the Zentralverband des Schornsteinfegerhandwerks (ZIV, Central Association of Chimney Sweeps) in version 1.0 of 01. August 2012, in version 2.0 of 13. February 2017 as well as version 3.0 from 02. July 2021. Please check with the manufacturer of your application program as to whether this interface is supported.
    • If the firmware update does not start under Windows 8.1 or Windows 10, a new bootloader must be installed on the measuring device once.
      A description and all necessary files can be found under the search term: Update-Kit / Bootloader
    • This software supports you in updating the software of your testo 330. Please read the instruction manual (german / english) for more information.
    • Update-Kit / Bootloader(V1.22, 1.24 MB)
      (testo 330 LL | testo 330i | testo 350 Control Unit + Analysis Box | testo 320)
      If the firmware update does not start under Windows 8.1 or Windows 10, a new bootloader must be installed on the measuring device once.