1. Home
    2. testovent 417

    testovent 417 - flow straightener

    0554 4172
    $ 127.50
    1
  1. Determining the volume flow by funnel measurement
  2. Reliable measurement results on swirl outlets
  3. Measurement of flow velocity
  4. Availability of different attachable measurement funnels
  5. Flow direction detection
  6. Hold function (to retain a reading)
  7. Flow straightener testovent 417.  Can be used with testo 417, and testo 440
    • testo-417-living-space-ventilation-02-2000x1500.jpg

      Determining the volume flow by funnel measurement

    Testo is the only manufacturer providing a solution the problem of measuring equally easily and accurately at swirl outlets. The innovative volume flow straightener testovent 417 supports you perfectly in flow velocity measurement at swirl outlets. The volume flow straightener allows redirection of the swirling flow to an almost uniform flow which can then be reliably recorded with a vane anemometer or vane probes. Measurements at swirl outlets without the volume flow straightener lead to lower or higher measurement results, depending on the flow direction in which the swirl impacts onto the vane.

    The volume flow straightener is used in combination with the optional measurement funnel testovent 417 and the optional vane anemometer testo 417 or the vane probes (Ø 100mm) of testo 435, testo 440 and testo 480.

    General technical data

    Weight
    20 oz. / 554 g
    Dimensions
    15 x 7 x 8 in. / 370 x 190 x 210 mm
    Operating temperature
    32° to 122 °F / 0 to +50 °C
    Housing
    PA66 GF30%

    Volume flow

    Measuring range
    0.1 to +118 CFM / 0.1 to +200 m³/h
    preferred 0.1 to +59 CFM / preferred 0.1 to +100 m³/h